Common Fault codes for LASER systems.
Candela MGY/MGL
Candela V-Beam
Candela PRO series
Cynosure apogee
Cynosure Elite/+
Cynosure Cynergy
Alma Soprano/Harmony
Zimmer Cryo 5/6
* Candela MGY/MGL and Gmax pro
Fault 3 , Is a shutter problem and has to be repaced
Fault 5.1 Fault 5.2 is a HVPS fault and needs to be replaced
Fault 6 is calibration issues where voltage exceeds required Joules
Fault 7 is water related
Fault 10 is delivery system issues (handpeice or fibre)
Fault 12 is internal energy errors
Fault 15 is external energy errors.
Fault 26 is software issues just restart device and it should work
*Candela V-Beam
Fault 4.1, 4.2, 5.1 is HVPS faults need to be replaced
Fault 11 is wavelength issues where the DYE-KIT most likely needs to be replaced
Fault 12.4 the Voltage drop on calibration is to high
​
*Cynosure Apogee/Elite
E01 - is either a fuse fault or pump
E16 - E17 -is a HVPS fault
E23 - external energy error (either fibre or lenses)
W54, W codes are warnings and need to be looked at but the system can still be used, it will become an E error eventualy
*Cynosure Cynergy
E33 PDL error, needs to be recalibrated and parts replaced
​
*Alma Soprano \ Harmony
E16 - either the device needs to be refilled with DI water or no pressure from pump to flow switch
E23 - (soprano) Either diode is blown or MEG module is blown
Simmer - ( harmony) either handpeice is broken or switching module
​
*Zimmer Cryo 5
F16 - is the temp sensor
F22 - The device is to warm, move device away from external heat source